Correct option is (b) False
Explanation: False, Signal processing for FHR determination can be based either on detecting the foetal heart valve motion or on detecting the heart wall motion in Ultrasound technique as The heart valve motion detection technique is based on the distinct ultrasound frequency shift produced by the fast opening and closing of the heart valves. The technique, however, requires that the ultrasound beam must be directed against the relatively small heart valves involving a longer search period and frequent repositioning of the transducer. Therefore, it is not preferred for continuous monitoring applications.