Correct choice is (a) True
The best explanation: Because kLa measurements involve monitoring levels of DO following a system perturbation, the results can be influenced by the response time (or “speed”) of a sensor making those determinations. Sensors with response times (τr) on the order of the first-order time constant of the mass transfer (1 / kLa) require special treatment of their data to correct for the time lag in readings introduced by the oxygen sensor.